Detailed description
 

Application: 
Water-cooled Weather Conditions Meter. To determine the color fastness and aging to sunlight, weather resistance, compound color fastness to sunlight-sweat of colored textile, leather, leatheroid, plastic, etc, by setting various parameters of irradiance, temperature, humidity, spray to simulate natural condition in lab. Be able to control irradiance real-time, automatically inspect and compensate irradiance, temperature, humidity.

Related Standards:

STANDARDS

ISO105-B02/B04/B06, 4892-2, 11341

IEC-60068-2-5

AATCC TM16, TM169

JIS L0843,GMW 3414

Note: this tester can conform to but not limit to all the standards above, for more standards conformance, please contact us.

Features:
l          Presetting various specifications, or customized programs, meet diverse range of standards, including AATCC, ISO, BS, JIS etc;
l          EviewTM touch screen, and build-in PLC (OMRONTM, Japan), real-time display the ongoing test’s graphic, be able to presetting irradiance, temperature, humidity, and plot dynamic curve
of them.
l          4500W water-cooled xenon arc lamp system, accurately simulate sunlight spectrum.
Irradiance auto- compensates system, control irradiance, temperature, and humidity timely through a closed loop.
l          Irradiation control, light irradiance sensor monitor, digital setting, automatic compensating system, stepless irradiation regulation.
l          Accurate chamber temperature control, irradiation temperature, and heater raise the temperature, air conditioning refrigeration system.
l          Ultrasonic humidification system makes spray water even and low noise. Saturated steam humidifying, industrial refrigerating machine dehumidifying, and close loop control the humidity.
l         Large amount of testing specimen, be able to hold as 3 times of specimen quantity as air-cooled machine in a test.
l          Custom program capability and storage for additional user-defined tests.
l           De-ionized water system to ensure purity of cooling water and spray water.
l           Each specimen holder has time recorder respectively. Ease to test different specimen in one test.
l          Multi-point inspection: malfunction alarm, self-diagnostic function, ease to maintenance.
l          Print the test result chart by stylus printer directly. No need support of PC.

Key Specification:

Power Supply                                              AC220V±5% 50Hz 9000W(Max)

Irradiation Source:                                           4500W Water-cooled Long Xenon Arc Lamp

Average Lamp Life                                       1000 hours/1500 hours ()

Exposure Area of Specimen                              2800

Chamber Temperature Range                       25℃85℃±2℃

BPT                                                           20℃120℃±2℃

BST:                                                             20℃130℃±1℃

Humidity         

Light Cycle                                                 10RH85RH ±5% RH

Dark Cycle                                                 10RH100RH ± 5% RH

Irradiance range

Wide Bandwidth Measurement:                      (35150)±1W/m2 @300400nm sensor 1#
Narrow Bandwidth Measurement:                   (0.8003.000)±0.02W/m2 @420nm sensor 2#
Full range(280-3000nm ) Measurement           Up to 2000W/m2@280~3000nm   sensor 3#
Sensor 1#, 2#, 3#, should be specified when place order.
Compressor Power:                                     1500W
DI water Container:                                       5000mL
Floor Space:                                                1050×3000mm (Min)

Instrument Dimension                                1200×670×1800mmL×W×H

Instrument Weight                                      400kg

Environmental Requirements (Ambient):
l        Temperature Range:            16-29℃ (60- 85oF)
l         Humidity Range:                0-85% RH, non-condensing. 
l         Altitude:                            Up to 2000 meters (6560 feet)
l         Storage:                            10-43℃ (50-110oF), 20-80% RH
l        Installation Category:           II
l         Pollution Degree:                2

The room should be ventilated properly with minimal drafts, air pollution, or dust. Do not install the instrument in a corrosion testing environment, or adjacent to corrosion testing equipment it is not advisable to install the instrument in such locations as a warehouse, boiler room, garage/factory are, paint room, or where doors to the outside environment are frequently opened causing changes in room temperature and humidity. Such locations may adversely affect instrument operation, longevity and test integrity (repeatability).

RELATED KEY WORDS:

Water-cooled Weather Conditions Meter,water cooled xenon meter,

Xenon test weatherometer, Xenon test fadeometer, Xenon weatherometer,

Xenon test weather-ometer,Xenon test fade-ometer ,Xenon test , Xenon meter,

Accelerated ageing test,  Weathering tester,  Weathering test , Light fastness,

Lightfastness

for more details, please contact us